Performance Qualification of Purified Water System

In a pharmaceutical company, we often need to conduct Performance Qualification. While performing Performance Qualification, we have to pay attention to many aspects, such as following and implementing each step of the process according to the protocol. Therefore, you can review each process and its steps listed below, which should be included in a protocol. Performance Qualification of Purified Water System describes the following content.

Performance Qualification of Purified Water System

Performance Qualification of Purified Water System Contents:

SERIAL NO.ITEM DESCRIPTION
1.0PROTOCOL APPROVAL
2.0OVERVIEW
2.1Objective
2.2Purpose
2.3Scope
2.4Responsibility
2.5Execution Team
3.0GENERAL CONSIDERATION/PREREQUISITE
4.0REVALIDATION CRITERIA:
5.0PERFORMANCE QUALIFICATION PROCEDURE
6.0SAMPLING FREQUENCY
7.0SPECIFICATION OF PURIFIED WATER
8.0PERFORMANCE QUALIFICATION OF FIRST PHASE
9.0DEFICIENCY AND CORRECTIVE ACTION(S) REPORTS(S)
10.0PERFORMANCE QUALIFICATION REPORT (PHASE-I)
10.1SUMMARY
10.2CONCLUSION
10.3FINAL REPORT APPROVAL (FIRST PHASE)
SERIAL NO.ITEM DESCRIPTION
11.0PERFORMANCE QUALIFICATION OF FIRST PHASE
12.0DEFICIENCY AND CORRECTIVE ACTION(S) REPORTS(S)
13.0PERFORMANCE QUALIFICATION REPORT (PHASE-II)
13.1SUMMARY
13.2CONCLUSION
13.3FINAL REPORT APPROVAL (SECOND PHASE)
14.0PERFORMANCE QUALIFICATION REPORT (PHASE-III)
15.0ABBREVIATIONS
16.0ANNEXURES
17.0DEFICIENCY AND CORRECTIVE ACTION(S) REPORTS(S)
18.0PERFORMANCE QUALIFICATION FINAL REPORT
18.1SUMMARY
18.2CONCLUSION
18.3FINAL REPORT APPROVAL (SECOND PHASE)
1.0PROTOCOL APPROVAL :

This Protocol for Performance Qualification of Purified Water System to be used for producing Purified Water.

FUNCTIONNAMEDEPARTMENTSIGNATUREDATE
PREPARED BY QUALITY ASSURANCE  
REVIEWED BY PROJECTS / ENGINEERING  
REVIEWED BY QUALITY CONTROL  
REVIEWED BY PRODUCTION  
APPROVED BY QUALITY ASSURANCE  
2.0OVERVIEW :

As an important Regulatory requirement, Performance Qualification becomes an important integral part of the qualification process.

2.1 OBJECTIVE:
The objective of the performance qualification of the water system is to establish and confirm that the generation and distribution system will operate within specified operating ranges. To develop and evaluate the system operation, sanitization, and maintenance procedures. To verify that the water produced and delivered to the points of use consistently meets the required quality specifications and acceptance criteria.
2.2 PURPOSE: 
The purpose of this protocol is to establish documentary evidence to ensure that the Purified water system received gives pre-determined quality of water consistently.
2.3 SCOPE: 
The scope of this protocol is to verify the Purified water system for the following boundaries and establish the operating ranges and procedures.

-To ensure that purified water meets the requirements of the USP and European Pharmacopoeia.
-To confirm the appropriateness of critical parameters and the operating ranges.
-To establish the reproducibility and reliability of the systems to establish Alert and Action levels.
-To validate and establish sanitization and regeneration procedures>
-To Evaluate the effects of seasonal changes.
2.4 RESPONSIBILITY: 
Write the persons along with the department’s name included in protocols along with their Responsibilities.
2.5 EXECUTION TEAM:
The execution team is responsible for the execution of this report. The execution team comprises of:
DEPARTMENTDESIGNATIONNAMESIGNATUREDATE
PROJECTS    
PRODUCTION    
QUALITY CONTROL    
QUALITY ASSURANCE    
3.0GENERAL CONSIDERATION/PREREQUISITE
3.1Approved Standard operating procedure of the equipment shall be available
3.2Approved analytical methods for testing the samples collected during the processing
3.3The installation and operational qualification of the equipment shall be successfully completed before the execution of the performance qualification.
3.4All the deficiencies and discrepancies related to the system that affect the product quality and corrective action taken shall be recorded in the appropriate section of the protocol.
3.5The analytical test results and other reports related to the system shall be attached to the performance qualification of the equipment and finally verified.
4.0REVALIDATION CRITERIA:
 The system shall be revalidated if
 There are any major changes that affect the performance of the equipment. After major changes in the components of the system
5.0PERFORMANCE QUALIFICATION PROCEDURE :

The performance Qualification of the purified water system shall be done in three phases. The purpose and sampling plan are described as below:

The initial phase (phase-one) typically begins only after the successful completion of operational qualification. The water generated during the various phases will be used for manufacturing as long as the water meets the specifications.

PHASE- I

During this phase, appropriate operating ranges will be established. Sampling should be done after each step in the treatment process and from each point of use. The incoming feed water will also be tested to verify compliance with the specifications. The detailed sampling plan is mentioned in section 07.0. At the end of this phase alert and action limits will be established. This alert and action limit will be used during phase two and beyond.

PHASE- II

This phase is to demonstrate that the system consistently operates within predetermined operating ranges and delivers water of the required quality (as specified) when operated in accordance with the SOPs. The sampling plan will be the same as phase-I.

PHASE- III

This phase will be continued for one year to verify the extended performance of system procedures, quantity, and quality of water despite possible seasonal variations of feed water.

At the end of this phase, the performance qualification is considered as completed and ongoing monitoring will be established as a continuous record of water quality.

PHASEPRIMARY OBJECTIVESTYPICAL DURATION
1Develop appropriate operating rangesDevelop and finalize operating, cleaning, and maintenance procedures. Demonstrate production and delivery of water of the required quality. 4 weeks
2Demonstrate consistent operation within established ranges. Demonstrate consistent production and delivery of water of the required quality.4 weeks
3Demonstrate extended performance Ensure that potential seasonal variations are evaluated and treatedOne year
6.0SAMPLING FREQUENCY :

The sampling points indicating sampling point number and their sampling frequencies are mentioned as below.

Sampling pointsSampling point NoFrequency
Phase -1Phase -2Phase -3
     
     
     
     
     
     

During the First phase and Second phase, samples shall be collected and tested daily for four weeks from all Sampling points. During the Third phase, feed water samples shall be collected and tested on a weekly basis, and purified water samples shall be collected and tested for every alternative user point so that all user points shall covered in two days (i.e. all odd no. points shall be collected one day and even no. points on next day).

7.0SPECIFICATION OF PURIFIED WATER :
TestLimits
Ph. Eur.USPIn-house
DescriptionA Clear, colorless, odorless, and tasteless liquidA Clear, colorless, and odorless liquidA Clear, colorless, and odorless liquid
pH5.0  –  7.05.0  –  7.0
Conductivity at 25°cNot More Than 1.3 ms/cmNot More Than 1.3 ms/cmNot More Than 1.3 ms/cm
Total Organic CarbonNMT 500 ppbNMT 500 ppbNMT 500 ppb
Nitrates0.2 ppm0.2 ppm
Heavy Metals0.1 ppm0.1 ppm
Total Aerobic Microbial count100 CFU /ml100 CFU /ml100 CFU /ml
Pathogens:
Escherichia coli Salmonella Species Pseudomonas-aruginosa Staphylococcus aureus
Should be Absent
Should be Absent
Should be Absent
Should be Absent
Should be Absent
Should be Absent
Should be Absent
Should be Absent
Should be Absent
Should be Absent
Should be Absent
Should be Absent

During the first phase, the operational parameters and the cleaning/sanitization procedures and

frequency will be developed. Sampling should be done daily after each step of the purification process

and at each point of use for four weeks.  The second phase of the system validation is to demonstrate that the system will consistently produce the desired water quality when operated in confirmation with the SOPs. The third phase of validation is to demonstrate the seasonal variation.

8.0: PERFORMANCE QUALIFICATION OF FIRST PHASE

8.1 Objective

The objective of this phase is to establish appropriate operating ranges and provide data for the development of cleaning and sanitization procedures, regeneration procedures, and frequencies.

8.2 Procedure

  • The Phase One Qualification will be carried out for 30 days.
  • The Purified Water system will be Operated/Regenerated/Sanitized as per respective Standard Operating Procedures.
  • A sample shall be collected from all the specified points (refer to sampling points schematic diagram as shown in annexure-1) at a specified frequency as shown in section 06.0
  •  Water samples shall be analyzed for the tests specified in section 07.0 as per the respective test procedures. 
  • Record the results of each sampling point in respective data sheets. 

8.3 Acceptance criteria

Water, which is collected from different locations in the generation system, shall meet the specified acceptance limits mentioned in section 7.0 Water, which is collected from a purified water storage tank and all user points, shall meet the specification mentioned in section 7.0.

In case of any failure investigation shall be carried out and rectification or modification shall be done if required.  In such case, the sampling and test period of Phase I study shall be extended for a further 2 weeks. During this period no production activity shall be carried out.

*Stress Operating Condition

a. Power failure test:

Verify the system performance in terms of water quality after power failure (5 minutes).

Collect water samples from all the sampling points from SP 01 to SP 10 before and after power failure and analyze for the tests specified in section 07.0.

b. Continuous circulation Loop testing

Keep water in continuous circulation in a loop and collect water samples from different sampling points preferably 2 user points and one from storage tank at frequent intervals of 12 hours, 24 hours, 48 hours, and 72 hours, and analyze for tests specified for respective points.

9.0DEFICIENCY AND CORRECTIVE ACTION(S) REPORT(S)
The following deficiency was identified and corrective actions taken in consultation with the validation team.
Description of deficiency :
 
 
 
 
 
Corrective action(s) taken :
 
 
 
 
 
 
 

Reviewed By:

Date

10.0PERFORMANCE QUALIFICATION REPORT (PHASE-I):
10.1SUMMARY : 
   
   
   
   
   
10.2CONCLUSION  : 
   
   
   
  
  

10.3 FINAL REPORT APPROVAL(FIRST PHASE):

Verified that all test cases required for phase one studies, and results data attached to this protocol or included in the qualification summary report are completed.

Signatures in the block below indicate that all tests in this phase of the Qualification Protocol have been reviewed and found to be acceptable and qualified for phase two studies.

NAMEDESIGNATIONDEPARTMENTSIGNATUREDATE
  PROJECTS / ENGINEERING  
  PRODUCTION  
  QUALITY  CONTROL  
  QUALITY ASSURANCE  
11.0PERFORMANCE QUALIFICATION OF SECOND PHASE :

11.1 Objective

The objective of this phase is to demonstrate that the system consistently operates within pre-determined operating ranges and produces water of the specified quality when operated in accordance with the SOPs.

11.2 Procedure

  • The phase two qualification will be carried out for 30 days.
  •  The Purified Water system will be Operated/Regenerated/Sanitized as per respective Standard Operating Procedures.
  • Sample shall be collected from all the specified points (refer to sampling points schematic diagram as shown in annexure-1) at a specified frequency as shown in section 06.0
  •  Water samples shall be analyzed for the tests specified in section 07.0 as per the respective test procedures. 
  •  Record the results of each sampling point in respective data sheets.

11.3 Acceptance criteria:

Water shall meet the specified acceptance limits established at the end of the First Phase Qualification.

In case of any failure w.r.t. CFU counts and Chemical test failure, sanitization, and regeneration frequency have to be reviewed. Any leakage or stagnancy of distribution system due to mis-operation or sampling error may result in microbial failure.  Hence an extensive investigation is required before revision for the requirement of frequency of sanitization.  Any microbial count, (though within limit), shall be identified and action shall be decided on the basis of type of organism and re-occurrence of the organism in the next sample(s) (if any).

12.0DEFICIENCY AND CORRECTIVE ACTION(S) REPORT(S)
The following deficiency was identified and corrective actions taken in consultation with the validation team.
Description of deficiency :
 
 
 
 
 
Corrective action(s) taken :
 
 
 
 
 
 
 

Reviewed By:

Date

13.0PERFORMANCE QUALIFICATION FINAL REPORT:
13.1SUMMARY :  
   
   
   
   
   
13.2CONCLUSION  : 
   
   
   
  
  

13.3 FINAL REPORT APPROVAL(SECOND PHASE):

 Verified that all test cases required for phase two studies, and results data attached to this protocol or included in the qualification summary report are completed.

Signatures in the block below indicate that all tests in this phase of Qualification Protocol have been reviewed and found to be acceptable and qualified for phase three studies.

NAMEDESIGNATIONDEPARTMENTSIGNATUREDATE
  PROJECTS / ENGINEERING  
  PRODUCTION  
  QUALITY  CONTROL  
  QUALITY ASSURANCE  
14.0PERFORMANCE QUALIFICATION OF THIRD PHASE :

14.1 Objective

The objective of this phase is to demonstrate that when operated for an extended time period, the system produces and delivers water of the required quality, despite possible seasonal variations of the feed water.

14.2 Procedure

  • The third phase of qualification will be carried out for one year.
  • The Purified Water system will be Operated/Regenerated/Sanitized as per respective Standard Operating Procedures.
  • A sample shall be collected from all the specified points (refer to sampling points schematic diagram as shown in annexure-1) at a specified frequency as shown in section 06.0
  • Water samples shall be analyzed for the tests specified in section 07.0 as per the respective test procedures. 
  • Record the results of each sampling point in respective data sheets.

14.3 Acceptance criteria

Water shall meet the specified acceptance limits established at the end of the Second Phase Qualification.

15.0Abbreviation:
 PWS: Purified water system
PW: Purified water
RW: Raw water
UV: Ultraviolet
SOP: Stander operating procedure
SAC: Strong acidic  cation
SBC: Strong basic cation
CFU: Colony Forming Unit

16.0  Annexure(s):

Sr.No.Annexure No.Title of Annexure
   
   
   
   
   
   
17.0DEFICIENCY AND CORRECTIVE ACTION(S) REPORT(S)
The following deficiency was identified and corrective actions taken in consultation with the validation team.
Description of deficiency :
 
 
 
 
 
Corrective action(s) taken :
 
 
 
 
 
 

Reviewed By:

Date

18.0PERFORMANCE QUALIFICATION FINAL REPORT:
18.1SUMMARY : 
   
   
   
   
   
18.2CONCLUSION  : 
   
   
   
  

18.3 FINAL REPORT APPROVAL(THIRD PHASE):

Verified that all test cases required for performance qualification studies, and results data attached to this protocol or included in the qualification summary report are completed.

Signatures in the block below indicate that all tests in this performance Qualification Protocol have been reviewed and found to be acceptable, considered to be safe, and qualified for regular use

NAMEDESIGNATIONDEPARTMENTSIGNATUREDATE
  PROJECTS / ENGINEERING  
  PRODUCTION  
  QUALITY  CONTROL  
  QUALITY ASSURANCE  

Leave a Comment