SOP on operation and calibration of autoclave

👤 Written by: Pankaj Sharma
Quality Control Specialist | Industry Experience

✓ Reviewed by: Pankaj Sharma - Quality Control Specialist
Reviewed for Quality Control accuracy, laboratory practices, analytical methods, and technical relevance

📅 Last Updated: October 31, 2026

An SOP on operation and calibration of autoclave tells staff how to run a steam sterilisation cycle safely. It also tells them how to keep the temperature, pressure and timer readings correct. Operation covers utility checks, air removal and leak tests. It also covers loading, the cycle, unloading and records. Calibration compares each instrument with a traceable standard and records the error. Calibration is not validation: IQ/OQ/PQ proves the cycle reaches sterilising conditions in the load. Indicator tape shows exposure only. It never shows sterility.

Infographic of the SOP on operation and calibration of autoclave showing the steam cycle, sensors and records
The autoclave cycle and the four instruments that the calibration SOP covers.

What Is the SOP on Operation and Calibration of Autoclave?

An autoclave is a closed steel chamber that sterilises materials with saturated steam under pressure. The SOP is the approved document that says how to run it, how to record each cycle and how to keep its instruments accurate.

Objective: to lay down the procedure for operating, recording and calibrating the autoclave.

Scope: every steam autoclave in microbiology, QC or production. These sterilise media, glassware and garments. They also sterilise equipment parts or waste.

Responsibility: the technical operator runs the cycle and fills the logbook.
The QC analyst or production officer reviews the printout.
Engineering staff calibrate and maintain the unit.
The QA officer approves records and deviations.
The head of department (HOD) is responsible for compliance.

How Does an Autoclave Kill Microorganisms?

Saturated steam gives up heat when it condenses on a cooler surface, and that heat kills microbes. The sterilising parameter is temperature. Pressure only follows it, because saturated steam at a given temperature always has the same pressure.

Eupry’s validation guide gives the common conditions: 121 °C at about 15 psi (about 1 bar) gauge. It also gives 132–134 °C at about 2 bar gauge. These are gauge pressures, read above atmospheric pressure. Hold time depends on the load and the validated cycle. Air left in the chamber lowers the steam temperature at that spot, which is why air removal matters.

What Are the Main Parts and Which Ones Need Calibration?

Description of autoclave parts lists the following.

  • Chamber: jacketed or single-wall stainless steel.
  • Steam supply or generator: plant steam, pure steam or an internal heater with DM water.
  • Air-removal system: a vent valve (gravity type) or a vacuum pump (pre-vacuum type).
  • Door: gasket and safety interlocks.
  • Safety valve: releases pressure above the set limit.
  • Steam trap and drain: removes condensate.
  • Controller: PLC or microprocessor with a printer.
  • Sensors and gauges: temperature probes, pressure transmitter and gauge, timer.

The calibration SOP covers the chamber temperature probe (Pt100, a platinum resistance sensor), the independent recording probe, the pressure gauge or transmitter and the timer. Annex 1 clause 8.50 expects you to record each heat cycle with accurate equipment. It also expects duplex probes on independent control and monitoring systems. Those systems must be able to abort or fail a cycle that does not match the validated parameters.

Gravity Displacement or Pre-Vacuum: Which Tests Apply?

FeatureGravity displacementPre-vacuum
Air removalSteam pushes air out through a ventVacuum and steam pulses remove air
Typical loadMedia, glassware, liquidsPorous, wrapped and hard goods
Typical cycle121 °C, 15–30 min for lab loads132–136 °C (134 °C nominal), 3–18 min by load
Air removal testNot applicableDaily
Vacuum leak testNot applicableWeekly

Exact values depend on the manufacturer and model; always follow your OEM manual and site SOP.

How Do You Operate an Autoclave? Step-by-Step Process Flow

Process flow for the SOP on operation and calibration of autoclave from utility check to logbook entry
Seven operating steps from utility check to logbook entry.
  1. Utility and gasket check. Read the last logbook entry, any open breakdown and the calibration label. Do not use the unit if the label is overdue. Check steam, water and compressed air supplies, the drain strainer and the door gasket. On a top-loading unit, keep the heating coil covered with DM water.
  2. Air removal and leak tests. On pre-vacuum units, run the air removal test (Bowie-Dick test) before the first load. Run the vacuum leak test on the weekly day. Do not load until both pass.
  3. Load per pattern chart. Follow the validated load pattern chart. Pack items dry, in wrapping that lets air out and steam in. Never sterilise sealed containers. Loosen caps on bottles.
  4. Run validated cycle. Close and lock the door. Select the validated cycle and note its unique cycle identifier. On a gravity unit, vent until steam flows steadily, then close the vent valve.
  5. Review printout. Check that the whole load reached temperature before the hold timer started. On pre-vacuum cycles, confirm the drain temperature record. Compare temperature, pressure and time with the cycle limits.
  6. Unload and inspect load. Wait until the gauge reads zero. Never open the door under pressure. Wear heat-resistant gloves and goggles, open the door away from your face and exhaust liquids slowly. Inspect every item for dryness, damaged packs and moisture. Remove any item that fails and report it.
  7. Record in logbook. Enter the cycle identifier, load, times, readings and signature (Annexure I). Mark the load as sterilised. The QA officer reviews the sterilisation record as part of batch release.

Annex 1 clauses 8.45 and 8.56–8.58 support these records and checks. Tape or a colour-change sticker shows the process occurred, nothing more.

Key Operating Parameters

ParameterTypical valueNote
Temperature, liquids and glassware121 °C minimum, hold 15–30 minValidated per load
Temperature, porous load132–136 °C, hold 3–18 min134 °C nominal
Chamber pressureAbout 1.0 bar (15 psi) gauge at 121 °C; about 2.0 bar (30 psi) gauge at 134 °CGauge pressure
Leak testPressure rise of 1.3 mbar/min or lessEupry value; manufacturer limit governs
F0Equivalent minutes at 121 °C, with z = 10 °COverkill cycles are commonly set at 12 min or more

How Do You Calibrate an Autoclave?

Calibration means comparing an instrument with a traceable reference and recording the error. Use reference standards traceable to national standards, ideally from an ISO/IEC 17025 accredited laboratory.

InstrumentMethodPoints
Chamber temperature probeCompare with a reference probe placed beside itWorking temperatures, such as 121 °C and 134 °C
Independent recorder probeCompare with the reference and with the controllerSame points
Pressure gauge or transmitterCompare with a reference gauge, in gauge pressureWorking pressures
TimerCompare with a calibrated stopwatch over a hold periodHold time
  1. The QA officer approves the calibration plan. Engineering staff isolate the empty autoclave.
  2. Record the as-found readings before any adjustment.
  3. Adjust only if a reading is outside the tolerance in the approved SOP. Record the as-left readings.
  4. Attach a label with the calibration date, due date and initials.
  5. Complete Annexure IV. The QA officer approves it.

Many plants calibrate every 6–12 months and after major repair. Regulations set no fixed interval. 21 CFR 211.68(a) asks for routine calibration, inspection or checking under a written program.

If an instrument is out of tolerance, open a deviation. Then assess every cycle run since the last good calibration. For example, if the chamber probe read higher than the reference, the real chamber temperature was lower than the printout showed. Those loads need review and possible re-sterilisation.

Preventive Maintenance

Check the door gasket and drain strainer every day. Test the vacuum weekly. Service the steam trap, safety valve and vent filter on the manufacturer’s schedule, and replace the gasket when it hardens or cracks. Trend cycle data and probe drift, because slow drift often shows before a failure.

Data Integrity of Autoclave Records

Keep the printout with the logbook entry. On PLC units, restrict user access and keep the audit trail on. Record corrections with date, reason and signature, never by overwriting.

  • Test before loading: run the air removal and leak tests first. A failed test costs minutes; a failed load costs a batch.
  • Read the printout, not the tape: the physical record is your evidence of the cycle.
  • Label sterilised loads: use the cycle identifier so anyone can trace the load.
  • Respect load patterns: a packed chamber traps air and gives wet loads.
  • Calibrate in gauge pressure: state the unit on the record so nobody confuses it with absolute pressure.
  • Never rush the door: wait until the gauge shows zero.

Annexures

Annexure I: Daily Usage Logbook

DateCycle identifierLoadStart/end timeHold temp/pressurePrintout attachedOperatorReviewed by

Annexure II: Air Removal Test (Bowie-Dick Type) Record

DateTest pack lotCycle temp/timeResult (uniform/not)OperatorQA officer

Annexure III: Weekly Leak Test Record

DateHold timePressure riseLimitPass/failOperator

Annexure IV: Calibration Record

InstrumentReference ID/certificateSet pointAs-foundAs-leftTolerancePass/failNext due

Annexure V: Cycle Printout Review

Cycle identifierWhole load at temperature before holdHold time metDrain temp recordedLoad dry and packs intactReviewer

Annexure VI: Breakdown/Deviation Record

DateDescriptionLoad statusImmediate actionRoot causeImpact on earlier cyclesQA approval

Annexure VII: Load Pattern Chart

Load typeContainer/packMaximum quantityPosition in chamberCycleRevalidation date

Frequently Asked Questions

What temperature and time are used in an autoclave?

Lab loads and media commonly run at 121 °C (about 15 psi gauge) for 15–30 minutes. Pre-vacuum cycles for porous loads run around 134 °C for shorter holds. The validated cycle for each load defines the final time, so follow the approved cycle rather than a general figure.

How often must an autoclave be calibrated?

Regulations require a written calibration program but give no fixed interval. Many plants calibrate temperature, pressure and timer every 6–12 months, and again after major repair. The site sets the interval by risk assessment and records it in the calibration SOP.

What is the difference between calibration and validation?

Calibration checks that the sensors, gauge and timer show correct values against a traceable standard. Validation (IQ/OQ/PQ) proves the cycle delivers sterilising conditions to every part of the load. A calibrated autoclave is not automatically validated, and a validated one needs calibrated instruments.

What is an air removal test?

It is a test cycle on an empty pre-vacuum chamber with a test pack. A uniform colour change shows the chamber had no trapped air and steam reached the pack. Annex 1 clause 8.61 expects it normally daily, or an air detector system as the alternative.

Is autoclave tape proof that a load is sterile?

No. Annex 1 clause 8.44 says such indicators show only that the sterilisation process occurred. Assurance comes from the validated cycle, the recorded temperature, time and pressure, and periodic biological indicators.

What should be recorded for each cycle?

Record the date, unique cycle identifier, load description, start and end times, hold temperature and pressure, operator and reviewer, and attach the printout. Annex 1 clause 8.45 requires a sterilisation record for each run, reviewed during batch certification. Failures need a deviation record.

References

  1. EU GMP Annex 1: Manufacture of Sterile Medicinal Products, European Commission (2022)
  2. Validation, Cleaning Validation, PIC/S, 2007
  3. Plant SOP (library)

Leave a Comment