SOP on Checking Balance Eccentric Accuracy

👤 Written by: Naresh Bhakar
Founder | Pharmaceutical Professional

✓ Reviewed by: Pankaj Sharma - Quality Control Specialist
Reviewed for Quality Control accuracy, laboratory practices, and technical relevance

📅 Last Updated: May 6, 2023
Stay Updated with Latest Pharma Trends
📢 Join Telegram 💬 Join WhatsApp

1.0 Objective: To lay down a procedure for checking balance eccentric accuracy.

2.0 Scope: The procedure applies to the balance eccentric accuracy in the production area.

3.0 Responsibility: Officer, Executive – Production Department
Manager – Production Department

4.0 Procedure:

4.1 Check the cleanliness of the weighing balance.
4.2 Calibrate and operate the weighing balance of 200 gm,3000 gm, 6 Kg, 30 Kg, and 60 Kg capacity weighing balance.
4.3 Select the standard weights for checking the eccentric accuracy for various balances based on the capacity as mentioned below:

Table for Balance Eccentric Accuracy

4.4 Place the standard weights at five different points of the platform (all four corners and middle) to check the balance’s eccentric accuracy and record the same in Annexure-I.
4.5 Displayed weight at five locations shall be approximately the same.
4.6 If the results exceed the acceptable (limit ± 0.2 % of the certified value of the standard weight), Put the ‘UNDER MAINTENANCE’ tag and inform the head of the department for necessary corrective action.
4.7 Ensure that the standard weights are certified by the weights and measures department.

4.8 Frequency: Monthly

Related Post: Weighing Balance Calibration in Pharmaceutical

5.0 Abbreviation

% – Percentage
g – Gram
kg – Kilogram
LC – Least count
NLT – Note Less Than
SOP – Standard Operating procedure
wt. – Weight

Annexure-1

Annexure for Balance Eccentric Accuracy

Leave a Comment