Home » Pharma SOPs » SOP on Checking Balance Eccentric Accuracy

SOP on Checking Balance Eccentric Accuracy

.0 Objective: To lay down a procedure for checking balance eccentric accuracy.

2.0 Scope: The procedure applies to the balance eccentric accuracy in the production area.

3.0 Responsibility: Officer, Executive – Production Department

Manager – Production Department

4.0 Procedure:

4.1 Check the cleanliness of the weighing balance.

4.2 Calibrate and operate the weighing balance of 200 gm,3000 gm, 6 Kg, 30 Kg, 60 Kg capacity weighing balance.

4.3 Select the standard weights for checking the eccentric accuracy for various balances based on the capacity as mentioned below:

Balance Eccentric Accuracy

4.4 Place the standard weights at five different points of the platform (all the four corners and middle) to check the balance’s eccentric accuracy and record the same in Annexure-I.

4.5 Displayed weight at five locations shall be approximately the same.

4.6 If the results exceed the acceptable (limit ± 0.2 % of the certified value of the standard weight), Put the ‘UNDER MAINTENANCE’ tag and inform the head of the department for necessary corrective action.

4.7 Ensure that the standard weights are certified by the weights and measures department.

4.8 Frequency: Monthly

Related Post: Weighing balance Calibration in Pharmaceutical

5.0 Abbreviation

% – percentage

g – Gram

kg – Kilogram

LC – Least count

NLT – Note Less Than

SOP – Standard Operating procedure

wt. – Weight

Annexure-1

Weighing Balance calibration FOR ECCENTRICITY

Leave a Comment